The Well@Home Monitor by Patient Care technologies Model 179-2150 Intended to be used as a clinical tool for remotely monitoring a patient's health status at home or in another healthcare facility on a spot check basis.

FDA Recall Z-2166-2010 · Posted 2010-08-03

Recall StatusTerminated
Product CodeDXN
Date Initiated2009-02-02
Date Terminated2010-08-03
Product Quantity62

Reason for Recall

Touch panel may become unresponsive.

Firm Action

Zoe Medical notified the 2 distributors by e-mail on January 26, 2009 and Feb 6, 2009 of the recall, requesting return of the units for correction.

Root Cause

Component change control

Distribution Pattern

Nationwide Distribution -- MA and MD.