Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 The FluidSmart Tube Sets are intended to be used with the FluidSmart Fluid Management System, which is intended for irrigation and fluid warming in laparoscopic procedures, and distention, fluid warming, and volume/deficit measurements in endoscopic procedures within gynecology, urology, and orthopedic disciplines.
FDA Recall Z-2166-2026 · Posted 2026-05-12
Advertisement
Recalling FirmStryker Corporation
Recall StatusOpen, Classified
Product CodeHIG
Date Initiated2026-04-09
Product Quantity10 tube sets
Reason for Recall
Due to nonconforming products being inadvertently distributed.
Firm Action
On 04/09/2026, the firm sent via USPS Certified Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers that Stryker identified that certain lots of FluidSmart tube sets were manufactured with an incorrect pressure relief valve specification. This defect may prevent the pressure relief valve from activating at the intended pressure in the event of a capital device over pressure failure. One lot (Lot B000576601, quantity 10) impacted by this nonconformance was inadvertently shipped to a customer.
Customers are instructed to:
1. Inform individuals within their organization who need to be aware of this action.
2. Check internal inventory to locate the product listed in the product affected grid.
3. If affected product is found, stop use immediately, segregate the product and call Stryker Customer Service at 1-800-624-4422
(Option 3) or email [email protected] to arrange for product return and issuance of credit or replacement (upon availability).
a. Remove the affected products, quarantine, and discontinue use of the identified product.
b. Please complete Business Reply Form on Attachment A on Page 2 and return it to [email protected].
4. If affected product is NOT found:
a. Please complete Business Reply Form on Attachment A on Page 2 and return it to [email protected].
5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Response is required.
For questions - [email protected] or 1-800-624-4422 (Option 3).
Root Cause
Process change control
Distribution Pattern
U.S. Nationwide distribution in the state of GA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.