Terminated

Esaote PA230E Transducer A portable and cart base ultrasound system for general and cardiovascular imaging

FDA Recall Z-2167-2014 · Posted 2014-08-08

Recalling FirmBiosound Esaote, Inc.
Recall StatusTerminated
Product CodeJOP
Date Initiated2014-06-26
Date Terminated2015-07-17
Product Quantity1,301 units

Reason for Recall

Using the PA230E transducer with a user-defined setting with the output power to 100% (Max) in 2D-TEI mode focusing at maximum depth, could cause the transducer surface temperature to exceed the maximum allowable limit.

Firm Action

On 06/26/2014, Esaote North America sent an advisory letter via FEDEX (with signature required) to each customer or consignee affected by this voluntary recall. The packet contained the advisory letter, including actions to be taken by the user, information about a pending software upgrade, and a business reply card to confirm they received the letter and have the equipment noted on the advisory letter. Customers with questions were instructed to call 800-428-7378. For questions regarding this recall call 1-317-813-6005

Root Cause

Software design

Distribution Pattern

Nationwide-Product was shipped to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI & WV. International**Product was also shipped to consignees in Canada