Endopath ETS-FLEX Endoscopic Articulating Linear Cutter, VASCULAR/THIN 35MM

FDA Recall Z-2174-2017 · Posted None

Recall StatusTerminated
Product CodeGDW
Date Initiated2017-03-31
Date Terminated2018-05-03
Product Quantity792 units

Reason for Recall

There is a risk that the pinion gear in the device could fail under extreme use cases.

Firm Action

Consignees were sent notification letters and were asked to complete the Business Reply Form (BRF).

Root Cause

Process control

Distribution Pattern

Worldwide distribution. The previously recalled products were distributed to the following countries: China, Japan, Singapore, and South Korea.