AxSYM Rubella IgG Reagent Pack; Abbott Laboratories, Abbott Park, IL 60064; list 3B23-20 The AxSYM Rubella IgG assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative and qualitative measurement of IgG antibodies to rubella virus in human serum or plasma (EDTA, heparin or sodium citrate) to aid in the determination of immune status to rubella.

FDA Recall Z-2175-2008 · Posted 2008-09-16

Recalling FirmAbbott Laboratories
Recall StatusTerminated
Product CodeLFX
Date Initiated2008-06-02
Date Terminated2008-11-05
Product Quantity33,485 reagent kits

Reason for Recall

An increase in complaints was noted for AxSYM Rubella IgG assay calibration failures due to error codes related to elevated Calibrator A (or Master Calibrator 1) rates too high.

Firm Action

Consignees were sent a Product Correction-Immediate Action Required letter on 6/2/08. The letter informed users of calibration failures due to elevates Calibrator A or Master Calibrator 1 rates when attempting to calibrate the AxSYM Rubella IgG assay and that they must clarify the Rubella IgG Calibrator A or the Rubella IgG Master Calibrator 1 and the Rubella IgG Negative Control by centrifugation at >/= 10,000 x g for 10 minutes prior to testing. Users were requested to return a customer reply letter acknowledging receipt of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution including USA, territory of Puerto Rico, and countries of Australia, Bahamas, Canada, Chile, Colombia, Costa Rica, Curacao, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Jamaica, Mexico, Peru, New Zealand, Panama, Singapore, South Korea, Thailand, Turks & Caicos and Uruguay.