Flexipet Manipulation pipette 300 micron. The device is used for the intracytoplasmic single sperm injection of oocytes.

FDA Recall Z-2177-2008 · Posted 2008-09-18

Recalling FirmCook Vascular Inc.
Recall StatusTerminated
Product CodeMQH
Date Initiated2008-03-02
Date Terminated2008-09-18
Product Quantity120 boxes

Reason for Recall

mislabeled - 300 micron flexipets labeled as 130 micron and vice versa

Firm Action

The recalling firm issued an Urgent product Recall letter on 3/3/08. The letter informed the customers of the problem and that the pouch is correctly labeled. Customers were instructed to return product and notify any sub accounts. Please contact Cook Vascular at 1-724-845-8621 for assistance.

Root Cause

Employee error

Distribution Pattern

The product was shipped to physicians in CA, FL, GA, IN, NJ, NM, NV, WA , and WI.