Terminated

Siemens Dimension(R) EZCR Flex(R) Reagent Cartridge. Catalog number DF270B. In vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma and urine on the Dimension Vista(R) System.

FDA Recall Z-2177-2011 · Posted 2011-05-10

Recall StatusTerminated
Product CodeJFY
Date Initiated2010-06-10
Date Terminated2013-03-19
Product Quantity860

Reason for Recall

Patient results may be falsely depressed or falsely elevated across the assay range.

Firm Action

Siemens issued an Urgent Field Safety Notice dated June 2010 to all Dimension(R) EZCR customers; to notify them that Siemens confirmed an issue which many cause EZCR results to be falsely depressed or falsely elevated across the assay range. The letter asks users to discontinue use of this lab test kit.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution.