Siemens Dimension(R) EZCR Flex(R) Reagent Cartridge. Catalog number DF270B. In vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma and urine on the Dimension Vista(R) System.
FDA Recall Z-2177-2011 · Posted 2011-05-10
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJFY
Date Initiated2010-06-10
Date Terminated2013-03-19
Product Quantity860
Reason for Recall
Patient results may be falsely depressed or falsely elevated across the assay range.
Firm Action
Siemens issued an Urgent Field Safety Notice dated June 2010 to all Dimension(R) EZCR customers; to notify them that Siemens confirmed an issue which many cause EZCR results to be falsely depressed or falsely elevated across the assay range. The letter asks users to discontinue use of this lab test kit.
Root Cause
Device Design
Distribution Pattern
Nationwide distribution.