Bard Access Systems, Inc., Gauderer Genie* II PEG (Percutaneous Endoscopic Gastrostomy) Safety System Ponsky "Pull" Placement Technique REF 000373, 20F, STERILE.

FDA Recall Z-2179-2009 · Posted 2009-09-15

Recall StatusTerminated
Product CodeKNT
Date Initiated2008-12-04
Date Terminated2011-11-09
Product Quantity177,614 units

Reason for Recall

Snare can detach in patient. The Polypectomy Snares and the Entake Peg System (which contain these snares) are being recalled because the looped portion of the snare can become detached and drop into the stomach or other internal canal during use, necessitating retrieval of the remaining piece.

Firm Action

Bard contacted all 1,883 worldwide customers who received the distributed suspect 467 product code lots, comprising 177, 614 units that were either not expired or were still available in the field. These will be returned by the customer to Bard Access Systems. 1-800-290-1689.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide, including USA, Belgium, Canada, Australia, and Japan.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.