KCI RotoProne Critical Care Therapy System buckles for securing patient to the surface during proning therapy, Model 209500. Indicated for treatment and prevention of pulmonary complications.
FDA Recall Z-2182-2011 · Posted 2011-05-10
Advertisement
Recalling FirmKCI USA, Inc.
Recall StatusTerminated
Product CodeIKZ
Date Initiated2011-02-18
Date Terminated2011-10-24
Product Quantity229 units
Reason for Recall
Buckles used to secure patient to surface during proning therapy can come open if not properly engaged by user.
Firm Action
Firm sent a letter with the "Warning: Improper Insertion of the RotoProne Therapy System Buckle Can Lead To Unintentional Bed Exit" to all consignees on 2/23/2011. The letter identified the affected product and the reason for the recall. Customers were instructed to read the attached product warning regarding the proper method to attach the RotoProne buckles and to post it with the device to ensure continued proper use of the buckles. If customers had questions regarding the proper use of the buckles, they were to contact KCI Customer Technical Service at 1-800-275-4524.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution - USA and Kuwait.