Terminated

IMPAX Cardiovascular (CV) Reporting Reporting tool used for structured reporting and structured report outputs (including XML style sheet customization and PDF)

FDA Recall Z-2184-2012 · Posted 2012-08-10

Recalling FirmAGFA Corp.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2012-07-27
Date Terminated2015-01-13
Product Quantity280

Reason for Recall

Content entered into the "Conclusions" free text box on the Report Writer screen was not represented on the final printed report.

Firm Action

AGFA Healthcare sent an URGENT FIELD SAFETY NOTICE dated July 27, 2012, to all affected customers. The letter described the safety alert and mitigation. Customer were asked to complete the Acknowledgment form and fax back to 864-421-1664 that the information was received and understood. For questions customers were instructed tol call 401-604-2180. For questions regarding this recall call 864-421-1754.

Root Cause

Software change control

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Canada

Related FDA 510(k) Records