Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System
FDA Recall Z-2188-2016 · Posted 2016-07-18
Advertisement
Recalling FirmToshiba American Medical Systems Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2016-02-24
Date Terminated2017-04-25
Product Quantity23
Reason for Recall
During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.
Firm Action
Toshiba Medical Systems (TAMS) planned action to bring product into Compliance:
1.TAMS will install the software and perform testing to ensure the software update was effective.
2. The customer notification letter which includes a statement that TAMS will, without charge, remedy the defect or bring the product into compliance.
.
If you have any questions regarding this, please feel free to contact your local Toshiba representative at (800) 521-1968.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.