Vital Port Vascular Access System Polysulfone Petite with detached 1.0mm catheter

FDA Recall Z-2191-2008 · Posted 2008-09-17

Recalling FirmCook Vascular Inc.
Recall StatusTerminated
Product CodeLJT
Date Initiated2008-04-24
Date Terminated2008-09-18
Product Quantity20 units

Reason for Recall

mislabeled - wrong size catheter was packaged in the box

Firm Action

The recalling firm issued an Urgent Product Recall letter dated 4/30/08 to inform their customers of the problem and the need to return the product. The letter also instructed the customers to notify sub accounts. Contact Cook Vascular at 1-724-845-8621 for assistance.

Root Cause

Packaging process control

Distribution Pattern

The products were shipped to distributors in Japan and Brazil.