Terminated

Vision One Laser System Model GA-0025020, Serial No: 10050, 10051, 10053. Intended for use in the treatment of ocular pathology.

FDA Recall Z-2191-2013 · Posted 2013-09-11

Recalling FirmLumenis, Inc.
Recall StatusTerminated
Product CodeGEX
Date Initiated2013-08-19
Date Terminated2013-12-19
Product Quantity3 units

Reason for Recall

Lumenis has initiated a recall on certain models of Vision One System due to a potential for unintended laser exposure to the user.

Firm Action

Customer's were notified via letter on 8/19/13. Service visits to consignees were scheduled with anticipated completion on 8/26/13 to replace the control board. Consignee monitoring was performed by use of return reply verification tracking cards returned by the Lumenis service engineer after completion of CPU board replacement.

Root Cause

Software design

Distribution Pattern

Distributed in the states of NJ, PA, and IL.