Open, Classified

Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.

FDA Recall Z-2192-2020 · Posted None

Recall StatusOpen, Classified
Product CodePDJ
Date Initiated2020-03-23
Product Quantity42,237 instruments

Reason for Recall

The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.

Firm Action

The recalling firm/manufacturer issued medical device correction letters dated 3/23/2020 on 3/23/2020 to their direct distributors requesting end user customer information. The recalling firm/manufacturer issued medical device correction letters to the end users dated 3/31/2020 via a third-party notification firm. The notification informed the end user of the issue and requested to be notified of the number of power supplies needed for replacement. The affected power supplies were to be discarded by the end user.

Root Cause

Device Design

Distribution Pattern

Distribution was made to NJ. There was no direct government/military distribution. Foreign distribution was made to Canada, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Denmark, Finland, France, Georgia, Germany, Guatemala, Hong Kong, Iceland, India, Italy, Ivory Coast, Japan, Korea, Madagascar, Malaysia, Mexico, Mongolia, Myanmar, The Netherlands, Norway, Oman, Pakistan, Peru, Philippines, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Timur, Trinidad & Tobago, United Kingdom, and Zimbabwe.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →