Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098, Tel: (800) 940-3240, Fax: (713) 529-7999 Software used to make on-screen (2-D, M-Mode and Doppler) measurements of left ventricle, right ventricle, etc.
FDA Recall Z-2193-2011 · Posted 2011-05-13
Advertisement
Recalling FirmDigisonics, Inc
Recall StatusTerminated
Product CodeLLZ
Date Initiated2009-02-06
Date Terminated2011-05-31
Product Quantity74 units
Reason for Recall
Left ventricle volume was inaccurate when the volumes were measured from images having different scales.
Firm Action
The firm, Digisonics, notified consignees/customers via e-mail entitled "Distribution of Notification of Biplane Error" dated February 6, 2009. The email described the product, problem and actions to be taken. The customers were instructed to immediately contact Digisonics if their site calculates biplane volumes-a utility will be provided which will run on their database and identify all studies in which biplane volumes are calculated. The customers were also instructed to share the following important information with all users of DigiView at their site. The email stated -until the software patch has been installed, biplane volumes should be calculated only from pairs of images having the same depth scale. Replacement DigiView/ERS software that prevents further occurrence of this problem will be provided if it appears their site could be impacted.
The Distribution of New Software Release was sent on March 6, 2009.
If you have any questions, please call Digisonics Customer Support at 800-344-4266
Root Cause
Other
Distribution Pattern
Worldwide distribution: USA (Nationwide) and countries of: Canada, Lebanon and New Zealand.