Prisma M100 PRE pump infusion set. Product number 107791. Each set is packaged in a plastic pouch, 4 pouches per carton box
FDA Recall Z-2199-2014 · Posted 2014-08-13
Advertisement
Recalling FirmGambro Renal Products, Incorporated
Recall StatusTerminated
Product CodeKDI
Date Initiated2014-07-18
Date Terminated2015-02-05
Product Quantity401,451
Reason for Recall
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Firm Action
Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution.