GEMINI TF Big Bore CT/PET System. The Philips GEMINI PET/CT Imaging Systems are a family of integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) systems suitable for a wide range of diagnostic applications.

FDA Recall Z-2199-2015 · Posted 2015-07-22

Recall StatusTerminated
Product CodeKPS
Date Initiated2015-07-01
Date Terminated2018-06-26
Product Quantity44 Units

Reason for Recall

Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.

Firm Action

On 7/6/2015 the firm sent Field Safety Notices to their customers.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide, Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, Israel, Italy, Japan, Lebanon, Netherlands, Portugal, Saudi Arabia, South Africa, Switzerland, Thailand, Venezuela & Vietnam.