LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
FDA Recall Z-2200-2026 · Posted 2026-05-14
Advertisement
Recalling FirmMICROVENTION INC.
Recall StatusOpen, Classified
Product CodeQCA
Date Initiated2026-04-06
Product Quantity38
Reason for Recall
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
Firm Action
On 4/6/2026, recall notices with TERUMO NEURO listed on notice top, were emailed to customers who were asked to do the following:
1) Stop using and quarantine affected devices.
2) Complete and return the acknowledgement form via email to
-Distributors should provide the recall notice to medical facilities or users to whom you have distributed affected products.
- Distributors should reconcile the completed forms from all customers and submit to recalling firm via email [email protected]
If you have questions about this recall contact firm at [email protected]
Root Cause
Mixed-up of materials/components
Distribution Pattern
International distribution to the country of China.