Carestream DRX-1 System Battery --- Made in U.S.A. for Carestream Health, Inc., Carestream Health, Inc., 150 Verona St, Rochester, NY 14608. --- The battery may be sold with the DRX-1 imaging device or as a stand-alone accessory.

FDA Recall Z-2202-2012 · Posted 2012-08-13

Recall StatusTerminated
Product CodeMQB
Date Initiated2012-06-12
Date Terminated2016-05-25
Product Quantity9971 units

Reason for Recall

Carestream DRX-1 System batteries manufactured before June 2012 due to two occurrences of lithium-ion battery overheating, which may cause burns.

Firm Action

Firm sent consignees "Urgent: Customer Notification and Field Corrective Action" letters on June 8, 2012. The letters described the issue and provided recommended actions. Service dealers will contact all affected consignees to evaluate the battery. U.S. customers with questions can contact Carestream Customer Care at 800-328-2910. Customers outside the United States should contact their local Carestream service number.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution, including Nationwide (USA).