VITROS Chemistry Products GLU Slides-Product Code:1707801. IVD VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems
FDA Recall Z-2202-2019 · Posted None
Advertisement
Recalling FirmOrtho-Clinical Diagnostics
Recall StatusTerminated
Product CodeCGA
Date Initiated2019-06-17
Date Terminated2020-11-09
Product Quantity1099 (5 pack x 60)=329700 slides
Reason for Recall
Potentially Biased Results using VITROS Chemistry Products GLU Slides
Firm Action
Ortho issued On 17 June 2019, a customer letter (Ref. CL2019-172) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to customers and distributors. The customers were notified of the issue and instructed: Discontinue using Lot 0048-0964-9287, discard any remaining inventory and switch to an alternate lot.
" If you do NOT have an alternate lot, consider the following alternate options:
o Use an alternate glucose method until your replacement order arrives.
o Consult with your medical director to determine if the bias observed and rate of occurrence when using the affected lot is acceptable for continued use, and if so, continue to use until your replacement arrives.
" Ortho will credit and/or replace your remaining inventory of Lot 0048-0964-9287. Please specify full and/or partial sales units remaining and indicate your preference of credit or replacement on your Confirmation of Receipt form.
" Complete the enclosed Confirmation of Receipt form no later than June 21, 2019.
" Post this notification by each system that processes VITROS GLU Slides or with your user documentation until you switch to an alternate lot.
" Please forward this notification if the product was distributed outside of your facility.
If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Root Cause
Process control
Distribution Pattern
US Nationwide Distribution - AL, AR, CA, CO, CT, FL, KS, KY, MA, MN, TN
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.