Biomet Vanguard CR femoral 57.5 mm left interlok, Sterile, REF 183022, Knee prosthesis component.

FDA Recall Z-2204-2010 · Posted 2010-08-10

Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2010-04-20
Date Terminated2011-01-19
Product Quantity5

Reason for Recall

A knee implant labeled as left was opened and the package actually contained a right knee implant.

Firm Action

On 4/13, 14 and 15 Biomet made phone calls to the consignees to inform them of the recall. The firm sent Urgent Medical Device Recall Notices dated 4/20/2010. The customers were instructed to immediately locate the recalled products and return them to the manufacturer.

Root Cause

Process control

Distribution Pattern

1 unit in US. 3 units in Biomet distributor inventory, 1 unit in Japan