Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001
FDA Recall Z-2206-2020 · Posted 2020-06-03
Advertisement
Recalling FirmMedtronic Inc.
Recall StatusTerminated
Product CodeLKK
Date Initiated2020-04-10
Date Terminated2021-04-27
Product Quantity24 Pumps
Reason for Recall
Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.
Firm Action
Medtronic initiated verbal communication with affected consignees on April 10, 2020 via phone call and provided affected consignees with an "Urgent Field Safety Notice" via email between the dates of 04/24/20 - 4/30/20.
. In addition to informing consignees about the recall, the firm asked consignees to take the following actions:
1. Immediately identify and quarantine all unused affected pumps in your inventory.
2. Return all unused affected pumps in your inventory to Medtronic. Your Medtronic Representative can assist you in the return and replacement of this pump as necessary.
3. Forward this notice to all those who need to be aware within your organization and to any organization where the affected pumps may have been transferred. Please maintain a copy of this notice in your records.
Root Cause
Process control
Distribution Pattern
No US Distribution. OUS: Netherlands, France, and Sweden
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.