ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system

FDA Recall Z-2206-2026 · Posted 2026-06-04

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2026-03-09

Reason for Recall

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Firm Action

Distribution of Customer Advisory Notice (CAN) to all affected customers began by Siemens on March 9, 2026.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US and Worldwide