Recall StatusTerminated
Date Initiated2019-07-19
Date Terminated2020-10-27
Product Quantity221 implants
Reason for Recall
The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.
Firm Action
Customers who received product/affected lot were to be contacted via Urgent Medical Device Recall Letter dated July 18, 2019 via USPS on July 19, 2019. In addition to informing customers about the recall, the recall notification asked customer to take the following actions:
1. Identify affected devices in your stock by confirming the information on the side label: NobelActive Internal NP 3.5x15mm, article 34128, lot 13080792.
2. Confirm the information on the top label as shown in figure 1 above.
3. Complete attached Customer Acknowledgment Form and return it to Nobel Biocare, via email to
[email protected] or fax 714-282-5073, within 5 days of receipt of this letter.
4. Return incorrectly labeled products (top label 3.5x11.5) to: Nobel Biocare
Attn: Returns PFA1917
22715 Savi Ranch Pkwy
Yorba Linda, CA 92887 .
5. Correctly labeled products (top label 3.5x15) do not need to be returned.
6. If you require any further information or support, please contact your customer support representative at 1-800-322- 5001 x.1527.
Root Cause
Labeling mix-ups
Distribution Pattern
US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI
OUS: None