Atlan A350XL. Model Number: 8621600. anesthesia workstation

FDA Recall Z-2208-2026 · Posted 2026-06-04

Recalling FirmDraeger, Inc.
Recall StatusOpen, Classified
Product CodeBSZ
Date Initiated2026-05-07
Product Quantity394 units

Reason for Recall

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Firm Action

Draeger notified consignees of the recall update and expansion on about 05/07/2026 via emailed letter titled Urgent Medical Device Correction Update. Consignees were made aware that due to continuous monitoring, Draeger has determined that an additional number of Atlan devices are affected by the issue and they are replacing ventilator motors. Consignees were instructed that a Draeger Service Representative will contact consignees to arrange a date for the ventilator motor assembly replacement, and until the corrective action is implemented, the device may continue to be used with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Consignees were also instructed to complete and return the provided Medical Device Correction Return Response Acknowledgment and Receipt Form and if products have been made available to third parties, please forward this Urgent Medical Device Correction notice to them.

Root Cause

Process change control

Distribution Pattern

US distribution in WI, AR, MN, WI, CO, MA, PA, AZ, FL, NY, SD, ND. International distribution to: Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chili, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Gabon, Germany, Ghana, Greece, Greenland, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Italy, Israel, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mauritius, Mexico, Morocco, Mozambique, Namibia, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Uzbekistan, Venezuela, Vietnam, Yemen