Terminated

Motorized Height Ceiling Suspensions (MCS) and motorized height movement (actuator) for X-ray systems. Usage: Ceiling suspension unit with an actuator to hold monitors and X-ray shields.

FDA Recall Z-2210-2010 · Posted 2010-08-11

Recall StatusTerminated
Product CodeIZI
Date Initiated2008-06-26
Date Terminated2010-12-02
Product Quantity1111 units

Reason for Recall

Monitor ceiling carriage may separate from the monitor ceiling suspension.

Firm Action

On 07/02/08, the firm sent a "Urgent - Device Correction" letter to all their consignees. The letter describes a product as X-Ray system with Monitor Ceiling Suspension and motorized height movement. The consignees are informed that the fixation bolts supporting the monitor ceiling carriage may come loose creating a situation which may allow the monitor ceiling carriage to separate from motorized height Monitor Ceiling Suspension and that would cause serious injury to patient and user. The firm has issued a mandatory Field Correction Action to correct the problem and the firm's technicians will inspect all ceiling suspension actuators for loose screws. If loose screws are found the actuator will be replaced. Consignees should call Philips Call Center at 800-722-9377 #5, #4, #1, and Reference FCO72200126.

Root Cause

Other

Distribution Pattern

United States and Canada.

Official Source

View FDA Recall →