Sabouraud Dextrose Agar w /Lecithin and Tween 20 Acumedia PN 91081 Product Usage: Used for the isolation of fungi from surfaces sanitized with quaternary ammonium compounds

FDA Recall Z-2212-2017 · Posted 2017-06-01

Recall StatusTerminated
Product CodeJSG
Date Initiated2016-08-08
Date Terminated2017-06-02
Product Quantity10 units

Reason for Recall

Storage temperature listed on the label was incorrect. Label stated 2-30C. Actual storage temperature is 2-8C.

Firm Action

On 8/8/2016 and 10/6/2016, notifications were sent to the affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, and information for the correction to product.

Root Cause

Error in labeling

Distribution Pattern

US in the state of IL