EnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton, single use, Rx, catalog/REF number 7G - REF ECP017G.Product Usage: Indicated for use in acquiring tissue for diagnosing breast abnormalities.

FDA Recall Z-2213-2019 · Posted None

Recall StatusTerminated
Product CodeKNW
Date Initiated2019-05-02
Date Terminated2022-03-03
Product Quantity125,292 units

Reason for Recall

There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product.

Firm Action

The recalling firm issued letters dated 4/19/2019 via FedEx on 4/19/2019 with proof of delivery to the U.S. customers. The letters sent to foreign countries vary depending on the format required for the country.

Root Cause

Component change control

Distribution Pattern

Worldwide distribution - US Nationwide distribution, including Puerto Rico. There was government and foreign distribution and no military distribution. Foreign distribution was made to Canada, Algeria, Australia, Austria, Belgium, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Martinique, Monaco, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Viet Nam.