Terminated

cobas Liat Analyzer

FDA Recall Z-2216-2016 · Posted None

Recall StatusTerminated
Product CodeOCC
Date Initiated2016-05-09
Date Terminated2021-03-24
Product Quantity2

Reason for Recall

Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.

Firm Action

Customers were notified of the recall via phone and email sent on 5/9/2016. They were instructed to take the following actions: Immediately discontinue use of the affected instruments until they have been reworked by a Roche Field Engineering Specialist. A Roche representative will contact you to schedule this service. Complete the attached fax form and fax it to 1-317-521-4815. File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center 1-800-800-5973 if you have questions.

Root Cause

Under Investigation by firm

Distribution Pattern

Distributed in NC and WA.