syngo Dynamics Version 9.5

FDA Recall Z-2219-2012 · Posted 2012-08-17

Recall StatusTerminated
Product CodeLLZ
Date Initiated2012-07-13
Date Terminated2013-07-15
Product Quantity30

Reason for Recall

If multiple uncompressed images/clips are sent to the syngo Dynamics server, around the same time, it is possible for an image from a patient to be placed in another patient's study.

Firm Action

On 07/13/2012, Siemens issued a "Customer Safety Advisory Notice" to affected consignees according to Update Instruction SY062/121S. The notice informs the customers of the potential issue with the identified affected product and provides details regarding the resolution. A software update to relieve the symptom of the issue will be distributed and installed by a Siemens Service Engineer on all affected systems under Update Instruction SY063/121S. A software update to correct this issue will be distributed and installed by a Siemens Service Engineer on all affected systems under Update Instruction SY064/121S.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution -- MA, GA, MO, IA, CA, PA, OR, CO, NY, NJ, WA, UT, OH, WI, MI, KY, and NC.