Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
FDA Recall Z-2223-2017 · Posted None
Advertisement
Recalling FirmRoche Diagnostics Corporation
Recall StatusTerminated
Product CodeJQP
Date Initiated2017-05-09
Date Terminated2018-02-02
Product Quantity3
Reason for Recall
a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a
potential risk to involved operators and can also lead to potential cross contamination
of samples in the affected racks.
Firm Action
Roche sent an Urgent Medical Device Correction letter dated May 9, 2017, to all affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Consignees were asked to complete the attached form and fax it to 877-906-8982 or email to [email protected]. Consignees with questions were instructed to contact the Roche Support Network Customer Support Center at 1-800-428-2336.
Root Cause
Device Design
Distribution Pattern
US distribution to WA only.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.