Oxford Gap Gauge SML 3/4mm. The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component.
FDA Recall Z-2224-2012 · Posted 2012-08-17
Advertisement
Recalling FirmBiomet U.K., Ltd.
Recall StatusTerminated
Product CodeIQO
Date Initiated2012-07-27
Date Terminated2013-02-15
Product Quantity119 units
Reason for Recall
A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model
remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an
increase of 1.0mm in the value of each dimension
Firm Action
Biomet sent a "FIELD SAFETY CORRECTIVE ACTION" letter dated July 24, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Response Form was attached for customers to complete and return. Contact the firm for questions regarding this notice.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.