3i Custom # 5753, OSCM2913, 2.9MMX 3.4MM X 13MM, OSSEOTITE CUSTOM, MICROMINI IMPLANT, LOT # 470258, Sterile. Implant Innovations, Inc. 4555 Riverside Drive, Palm Beach Gardens, FL 33410. Custom temporary cover screws are for use in maintaining the soft tissue during the healing process of a dental restoration.
FDA Recall Z-2227-2009 · Posted 2009-09-22
Advertisement
Recalling FirmBiomet 3i, LLC
Recall StatusTerminated
Product CodeNDA
Date Initiated2006-12-11
Date Terminated2009-10-01
Product Quantity24
Reason for Recall
Healing Abutment was not made within required manufacturing specification. The product did not have the relief behind the thread, which does not allow proper seating.
Firm Action
Each of the US customers was contacted by phone. Following the phone contacts, Biomet 3i performed a follow- up of each call with a faxed letter. Each customer was instructed to check their respective inventory for the affected product and corresponding lot number. They were then requested to fax back a response form to Biomet 3i Regulatory Services to provide feedback concerning their inventory. They were also instructed to return the affected product to Biomet 3i. Contact Biomet 3i, LLC at 1.561.776.6706, if there are any questions.
Root Cause
Nonconforming Material/Component
Distribution Pattern
USA and Italy
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.