Terminated

Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;

FDA Recall Z-2228-2017 · Posted 2017-05-24

Recalling FirmConvaTec, Inc
Recall StatusTerminated
Product CodeBYG
Date Initiated2017-05-10
Date Terminated2021-08-27
Product Quantity816,709

Reason for Recall

Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.

Firm Action

Recall Letters were sent to consignees on May 10, 2017. The letter requested that they immediately stop distribution and quarantine all affected lots. The letter also requested a sub-recall if the product had been further distributed. Also attached to the letter is a Recall Response Form to be returned.

Root Cause

Process control

Distribution Pattern

CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑