Siemens brand Ultrasound Transducer 17L5HD used with ACUSON Sequoia Ultrasound System; Model Number(s): 10035724, used with System Models: 08245875, 08246951, 08255412, 08255413, 08267697, 08269627, 10038241, 10038242, 10040724, 10040725, 10041008; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Division, Mountain View, CA The 17L5 transducer is intended for breast and small body parts imaging.

FDA Recall Z-2230-2008 · Posted 2008-09-11

Recall StatusTerminated
Product CodeIYO
Date Initiated2008-06-06
Date Terminated2010-12-21
Product Quantity388 units

Reason for Recall

The product assembly can cause a failure of the transducer which results in a double and/or overlapped image which may ultimately result in misleading or false information, inability to accurately diagnose, incorrect positioning/locating/insertion of the biopsy needle and/or an interruption during the biopsy procedure.

Firm Action

On 6/6/08, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the affected product be returned. Affected transducers will be replaced. For assistance, contact Siemens Medical Solutions at 1-650-694-5993.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution: USA, AUSTRALIA, AUSTRIA, CANADA, CHINA, DENMARK, FRANCE, GERMANY, INDIA, ITALY, JAPAN, KOREA, MALAYSIA, MEXICO, NORWAY, PORTUGAL, RUSSIAN FEDERATION, SINGAPORE, SWEDEN, TAIWAN, and THAILAND.