VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease.

FDA Recall Z-2232-2014 · Posted 2014-08-14

Recalling FirmBioMerieux SA
Recall StatusTerminated
Product CodeDAP
Date Initiated2013-04-17
Date Terminated2014-10-08
Product Quantity2160 units

Reason for Recall

Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.

Firm Action

A recall letter was issued on 4/17/2013 to US customers via FED Ex mailing. The letter included a fax back acknowledgment form. A recall notice was also sent via email to bioMerieux Subsidiaries and international distributors on 4/11/2013.

Root Cause

Under Investigation by firm

Distribution Pattern

AL, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MA, MI, MN, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WA, WV, WI, WY