StatSampler 200 uL, Li Heparin (micro blood collector) Ref: SS2H Packed in Catalog Numbers: SS2H (I vial), SS2H-06 (6 kits/case), CH2C-I0 (I0 vials/kit) Intended for collection and processing of capillary blood to prepare plasma/serum.

FDA Recall Z-2234-2008 · Posted 2008-08-23

Recall StatusTerminated
Product CodeJKA
Date Initiated2008-06-20
Date Terminated2012-05-30
Product Quantity171 vials

Reason for Recall

EDTA inadvertently mixed with lithium heparin solution contained in capillary tubes.

Firm Action

Iris Sample Processing notified customers by telephone followed by an Urgent: Product Recall letter dated 6/19/08. Users are requested to return inventory, to not use product from Lot Number 7296, and to repeat tests where the recalled tubes were used. The firm identified an additional lot : 8031 was effected and recalled the lot to two consignees.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Worldwide Distribution --- USA including states of CA, CO, FL, KY, KS, GA, NC, NJ, OR, PA, and countries of Canada, Hong Kong, UK.