S&W Electrodes for Defibrillation, 10 Defibrillation Electrodes, Reorder #90101, S & W Healthcare Corp, 1525 Flight Path Drive Brooksville, Florida 34604 The device is non-sterile and single use. For use on adults and children over eight years old for external defibrillation, pacing, monitoring, and cardioversion.

FDA Recall Z-2234-2009 · Posted 2009-09-24

Recalling FirmLeonhard Lang Gmbh
Recall StatusTerminated
Product CodeMKJ
Date Initiated2009-07-10
Date Terminated2011-12-17
Product Quantity3044

Reason for Recall

A potential risk was identified associated with the connectors of the defibrillation electrodes DF20. It was found that some of the connectors do not slip easily into the sockets of the defibrillator or an extension cord of the defibrillator. In such cases increased force or some wiggling is required to insert the connector fully. A potential risk for patients exist, if (a) such a connector (b)

Firm Action

A Leonhard Lang letter dated June 9, 2009 was sent June 6, 2009 to its customer. The letter identified the problem and the product. The letter instructed the consignee to stop using or distributing the product and to quarantine them. It also requested them to report all quantities of product that they had in stock using the attached form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US - FL Foreign: Austria, Germany, Romania, France, Dubai, Greece, Italy, South Korea, United Kingdom, Hungary, Latvia, Saudi Arabia and Lebanon