2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and Sternotomy. They are intended to be used with the Stryker CORE" system.
FDA Recall Z-2238-2016 · Posted 2016-07-19
Advertisement
Recalling FirmStryker Instruments Div. of Stryker Corporation
Recall StatusTerminated
Product CodeHBE
Date Initiated2016-06-24
Date Terminated2017-03-23
Product Quantity3,163
Reason for Recall
A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be
more susceptible to breakages.
Risk to Health:
Injury to critical soft tissue requiring surgical intervention may occur.
Firm Action
Stryker sent an Urgent Medical Device Recall Notification letter dated June 24, 2016, to the affected customers via courier, and to Sales Representatives and international affiliates via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments at 269-389-2921.
Root Cause
Process design
Distribution Pattern
Worldwide Distribution - US (Nationwide) Netherlands, Canada, Japan, Switzerland, and United Kingdom
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.