PoleStar FlexiCoil band D/E, revisions A and B, Catalog # 9733063, Usage: Low field MRI device used to gather cranial images.

FDA Recall Z-2240-2010 · Posted 2010-08-19

Recall StatusTerminated
Product CodeLNH
Date Initiated2010-07-01
Date Terminated2010-11-09
Product Quantity560 units

Reason for Recall

Electronic box on headstrap may overheat, causing burn to patient.

Firm Action

Customers were notified by letter on 07/07/2010, advising them of the overheating problem and alerting them that a service rep will be visiting to inspect and upgrade any affected FlexiCoils. Questions were referred to 800-595-9709.