Terminated

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part: "***Neusoft***NeuViz16 MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED FOR NEUSOFT MEDICAL SYSTEMS CO., LTD***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD***ADDRESS: NEUSOFT PARK, HUNNAN INDUSTRIAL AREA, SHENYANG***CHINA***".

FDA Recall Z-2241-2011 · Posted 2011-05-23

Recall StatusTerminated
Product CodeJAK
Date Initiated2011-02-09
Date Terminated2011-08-24
Product Quantity9 units distributed in US

Reason for Recall

Dimension measurement displays on a combined image were found inaccurate after the clinician merged multiple image series into one image.

Firm Action

Neusoft Medical Systems sent an Urgent Device Correction Notification letter dated February 9, 2011, via certified mail to their distributor NEUISYS. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to discontinue use of combined images. Instead, use MPR Batch for diagnostic reviews until the planned upgrade is provided to the product. Neusoft Medical System and Philips Medical System released the mandatory Field Change Order (FCO) by March 31, 2011. All nine affected systems will be updated with aforementioned kits in six months after the release. Consignees were instructed to contact Service Support Department by e-mail to [email protected]

Root Cause

Other

Distribution Pattern

Nationwide Distribution including CA, GA, LA, NC, OH, TN and TX.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.