Terminated

Vanguard 360 Femoral Boss Reamer, Label: REF 32-488474 VANGUARD 360 1NSTRUMENT FEMORAL BOSS REAMER NON-STERILE, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE P.O. BOX 587 WARSAW, IN 46581 USA

FDA Recall Z-2242-2014 · Posted 2014-08-15

Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2014-07-01
Date Terminated2015-10-30
Product Quantity461

Reason for Recall

The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.

Firm Action

Urgent Medical Device Recall Notices dated July 15, 2014 were mailed with verification of delivery via FedEx. Response forms were provided to all consignees in order for Biomet to track returns. The letters informed the consignees to return the devices to Biomet and conduct a sub-recall sending the recall letter to all customers.

Root Cause

Process change control

Distribution Pattern

California, IA, FL, MD, UT, SC, AL, IL, CO, MA, KS, GA, WA, NV, AR, NH, OK,Pennsylvania, New York, Texas, Virginia, New Jersey, North Carolina, Kentucky, Michigan, Ohio, Wisconsin, Louisiana, Arizona, Indiana, South Dakota, Missouri & Montana. In addition to Korea, B.V., Japan, Puerto Rico, UK, Chile, New Zealand, Australia, & Canada