smith&nephew BHR SQUARE HEADED NAIL, REF CATALOG NUMBER 999908. For use during orthopedic surgery.
FDA Recall Z-2242-2021 · Posted None
Advertisement
Recalling FirmSmith & Nephew Orthopaedics Ltd. (Aurora)
Recall StatusTerminated
Product CodeNXT
Date Initiated2021-07-14
Date Terminated2024-05-13
Product Quantity47 units
Reason for Recall
The nail head may become detached during surgery.
Firm Action
The firm notified its consignee by letter on 07/14/2021. The letter explained the issue and the hazard and requested the return of the products.
Root Cause
Component change control
Distribution Pattern
Worldwide distribution - US Nationwide distribution and the countries GB, NO, CA, AU & NZ.