MERITMEDICAL Prelude Short Sheath Introducer, REF: PSS-6F-4-018MT, 21G 6F(2.0 mm) 4 cm, Max Guide Wire 0.018" (0.46 mm), STERILE EO, Single Use, Rx Only, CE2797, UDI: (01)00884450015391

FDA Recall Z-2243-2021 · Posted None

Recall StatusTerminated
Product CodeDYB
Date Initiated2021-07-21
Date Terminated2023-03-13
Product Quantity373 UNITS

Reason for Recall

There is a potential that a package sheath contains an incorrect dilator.

Firm Action

Merit Medical issued an "Urgent Medical Device Recall Notice" on July 12, 2021 via FedEx to all affected consignees. In addition to informing consignees about the recall, Merit Medical request consignees take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this recall. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan, and email the completed CRF to Customer Service at [email protected] within 10 business days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. 6. Adverse events experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online at www.fda.gov/medwatch or by phone at 1-800-332-1088. 7. If you have any questions concerning this communication, please dont hesitate to contact your Merit Sales Representative or Merit Customer Service at (801) 208-4381 | Hours: 6 am to 6 pm MDT | Mon-Fri.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of ID, HI, TX, CA, AZ, MN, UT, KS, MO and the countries of Republic of Korea, Netherlands, Germany, Cyprus, Italy, Sweden, Taiwan, and Spain.