WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01

FDA Recall Z-2243-2024 · Posted 2024-07-01

Recall StatusOpen, Classified
Product CodeHRX
Date Initiated2024-05-24
Product Quantity1700 units

Reason for Recall

Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery

Firm Action

Smith & Nephew Issued Urgent Field Safety Notice (FSN) - MEDICAL DEVICE RECALL WOM reference no.: 2024-0001 via email on 5/24/23. Letter states reason fo recall, health risk and action to take; lnform individuals within your organization who need to be aware of this recall. - Check all stock areas and/or operating room storage to determine if any tube sets with lol numbers from Annex 'l are at your facility, - Dispose of all affected tube sets properly. Confirm this on the customer response form. Return the form to [email protected]. Please send any questions to [email protected]

Root Cause

Process control

Distribution Pattern

US Nationwide distribution.