BacT/ALERT - FA Reagent Bottle The BacT/ALERT Microbial Detection System is used to determine if microorganisms are present in blood or other normally sterile body fluid samples taken from a patient suspected of having bacteremial/fungemia. The BacT/ALERT System and culture bottles provide both a microbial detection system and a culture media with suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and other normally sterile body fluid infections. An inoculated bottle is placed into the instrument where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT FA bottle. BacT/ALERT FA Culture Bottles are used with the BacT/ALERT Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and facultative anaerobic microorganisms (bacterial and fungi) from blood and other normally sterile body fluids.

FDA Recall Z-2244-2011 · Posted 2011-05-24

Recalling FirmbioMerieux, Inc.
Recall StatusTerminated
Product CodeMDB
Date Initiated2011-04-13
Date Terminated2012-04-03
Product Quantity327,700 bottles

Reason for Recall

The charcoal suspension in a portion of the lot does not meet the intended density.

Firm Action

BioMerieux sent an Urgent Product Removal Notice dated April 13, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Evaluate positive results adhering to their institution's protocols for determining false versus true positives. Negative bottles should be checked by smear and/or subculture at some point prior to discarding as negative. Any remaining inventory of BacT/ALERT FA culture bottle lot 1027481 should be destroyed in accordance with their facility's policies and procedures. An Acknowledgement and Product Replacement Form was included to be completed and returned to bioMerieux or credit or replacement of unused bottles.

Root Cause

Other

Distribution Pattern

Worldwide Distribution - USA (nationwide) and the countries of China, Tokyo, South Korea, Canada, Sweden, Austria, Poland, and Belgium