Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.

FDA Recall Z-2244-2012 · Posted 2012-08-22

Recall StatusTerminated
Product CodeOXO
Date Initiated2012-07-11
Date Terminated2016-06-06
Product Quantity17 units

Reason for Recall

A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.

Firm Action

Medtronic Navigation issued an "Urgent Field Safety Notice" to consignees dated 7/11/12. The problem was described and recommended actions were provided. For each affected O-arm Imaging System, a Medtronic service representative will visit the site and replace the 500VDC relay on the system and test the O-arm Imaging System.

Root Cause

Component change control

Distribution Pattern

Worldwide Distribution - USA, including the states of AL, CO, FL, IA, MI, IA, NC, NE, NY and the countries of South Africa, Italy, and Spain.