Magnetic Resonance Imaging Diagnostic Device accessory - Product Usage: The chest pneumograph (also referred to as the pneumatic bellows, respiration bellows, or respiration sensor) monitors bellows-derived respiration by detecting abdominal or chest wall motion. The chest pneumograph is an accessory to a MRI patient monitoring system or MRI gating solution. The Chest Pneumograph is indicated for use by healthcare professionals with the Wireless Sp02 (wSp02) Module or Toshiba KSB gating unit when bellows-derived respiration sensing is needed during MRI scans for gating.
FDA Recall Z-2245-2020 · Posted None
Advertisement
Recalling FirmPhilips North America, LLC
Recall StatusTerminated
Product CodeMWI
Date Initiated2020-04-13
Date Terminated2023-04-14
Product Quantity1075 Units
Reason for Recall
The labeling of the Philips Chest Pneumograph does not include a statement indicating that the product contains natural rubber latex.
Firm Action
In March of 2020 Philips Healthcare sent out an Urgent Medical Device Corrective Action Notice identifying the problem, under what circumstances it can occur, the actions that should be taken by the customer/user to prevent risks for patients/users, and the actions planned by Philips to correct the problem.
PRODUCTS: All Pneumograph, chest, NM, 3160 with part #94023 are affected.
ACTIONS TO BE TAKEN: Customers/users are asked to review the Notification letter and share with all members of staff who utilize this product. Customers/Users should also retain a copy with the equipment Instructions for Use. The chest pneumograph can continue to be used for monitoring as directed by a physician.
1.Identify all affected chest pneumographs in your facility using the guidelines provided in the HOW TO IDENTIFY AFFECTED PRODUCTS section above.Do not return any product to Philips.
2. Complete and sign the reply form provided on the last page of this letter.
3. Send the completed and signed reply form to Philips via the contact information located on the form.
4. Upon receipt of the reply form, Philips will send you one kit per chest pneumograph in your possession along with instructions for labeling each chest pneumograph.
5. Upon receipt of the kit, follow the included instructions to label each chest pneumograph in your possession. The label to be added identifies the product as containing latex.
ACTIONS PLANNED BY PHILIPS:
Upon receipt of the completed and signed reply form, Philips will provide one kit per chest pneumograph in each customers possession for the purposes of labeling each affected chest pneumograph. This kit will be provided free of charge.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.