Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set
FDA Recall Z-2246-2018 · Posted None
Advertisement
Recalling FirmSynthes (USA) Products LLC
Recall StatusTerminated
Product CodeFSM
Date Initiated2018-03-05
Date Terminated2020-02-28
Product Quantity4
Reason for Recall
Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.
Firm Action
On March 8, 2018, the recalling firm sent Urgent Product Recall letters to its customers to inform them of the recall. Customers were advised to take the following actions: 1. Immediately review their inventory to identify and quarantine all affected products listed above in a manner that ensures the affected products will not be used.
2. Complete the Verification Section and send a copy to the recalling firm. 3. Forward this notice to anyone in the facility that needs to be informed.
4. If any of the affected products have been forwarded to another facility, contact that facility to arrange return.
Customers with questions were advised to call 610-719-5450, 8AM-5PM (EST), Monday through Friday, or contact their DePuy Synthes sales consultant.
Root Cause
Process control
Distribution Pattern
Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.