Terminated

SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.

FDA Recall Z-2250-2017 · Posted None

Recalling FirmCircadiance LLC
Recall StatusTerminated
Product CodeNPE
Date Initiated2017-05-01
Date Terminated2018-02-05
Product Quantity1,242 units

Reason for Recall

Circadiance has determined that it is possible for certain Smart Monitor 2PS/PSL monitors to exhibit intermittent operation of the nurse call feature. The firm has updated the device to increase the "Nurse Call" feature to ensure monitor alarms are transmitted to compatible nurse call systems.

Firm Action

Circadiance mailed a Medical Device Removal SmartMonitor 2 PS/PSL, dated 04/28/2017, to affected customers to inform them of the issue, the actions to be taken and the actions the firm plans to take to correct the issue. Customers are told to test the functionality of the system prior to each use; and, complete the Business Reply Form. The form can be returned via 412-202-4583 or emailed to [email protected]. Customers are to contact Circadiance Customer Service at 724-858-2837 to receive an Return Material Authorization number; or, 1-888-825-9640 to repair their unit free of charge, or for questions or support.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Distribution.